Zum Inhalt springen

Nur-Lesen-Fassung: Das Forschungsinformationssystem (FIS) ist aus Sicherheitsgründen vorübergehend durch diese Nur-Lesen-Fassung ersetzt. Alle öffentlichen Inhalte und Volltexte sind weiterhin abrufbar; Anmeldung, Bearbeitung, Einreichung und Statistiken stehen nicht zur Verfügung. Bei Fragen: fis@uni-bamberg.de

Start › Publikationen › Fast 3D YOLOv3 based standard plane regression of vertebral bodies in intra-operative CBCT volumes

Publikation

Fast 3D YOLOv3 based standard plane regression of vertebral bodies in intra-operative CBCT volumes

Dörrich, Sebastian; Kordon, Florian; Denzinger, Felix; u. a. (2023): Fast 3D YOLOv3 based standard plane regression of vertebral bodies in intra-operative CBCT volumes, in: Bamberg: Otto-Friedrich-Universität, S. 1–15.

Details

Fakultät/Lehrstuhl
Autorin(nen)/Autor(en)
Kordon, Florian
Denzinger, Felix
El Barbari, Jan S.
Privalov, Maxim
Vetter, Sven Y.
Maier, Andreas
Kunze, Holger
Verlagsangabe
Otto-Friedrich-Universität : Bamberg
Erscheinungsjahr
2023
Quelle/Weitere Ausgabe
Journal of medical imaging : JMI, 10 (2023), 3, S. 1-15 . - ISSN: 2329-4310
Jahr der Erstpublikation
2023
Sprache(n)
Englisch
Lizenz
URN
Abstract

Purpose

Mobile C-arm systems represent the standard imaging devices within the field of spine surgery. In addition to 2D imaging, they allow for 3D scans while preserving unrestricted patient access. For viewing, the acquired volumes are adjusted such that their anatomical standard planes align with the axes of the viewing modality. This difficult and time-consuming step is currently performed manually by the leading surgeon. This process is automatized within this work to improve the usability of C-arm systems. Thereby, the spinal region consisting of multiple vertebrae and the standard planes of all vertebrae being of interest to the surgeon need to be taken into account.

Approach

An object detection algorithm based on the you only look once version 3 architecture, adapted to 3D inputs, is compared with a segmentation-based approach employing a 3D U-Net. Both algorithms are trained on a dataset of 440 and tested on 218 spinal volumes.

Results

Although the detection-based algorithm is slightly inferior concerning the detection (91% versus 97% accuracy), localization (1.26 mm versus 0.74 mm error) and alignment accuracy (5.00 deg versus 4.73 deg error), it outperforms the segmentation-based one in terms of speed (5 s versus 38 s).

Conclusions

Both algorithms show similar good results. However, the speed gain of the detection-based algorithm, resulting in a run time of 5 s, makes it more suitable for usage in an intra-operative scenario.

GND-Schlagwort(e)
Bilderzeugung; Norm <Normung>; 3D-Technologie; Objekterkennung; Bildsegmentierung; VOXEL; Computertomograf; Operation; Anatomie; Wirbelsäule; Algorithmus; Mustererkennung
Freie Schlagwort(e)
3D imaging standards; Object detection; Image segmentation; Cone beam computed tomography; Surgery; Detection and tracking algorithms; Education and training; Anatomy; Spine; Voxels
DDC-Klassifikation
004 Computer science
610 Medizin, Gesundheit
RVK-Klassifikation